Section 15: Calibration  Series 3700 System Switch/Multimeter Reference Manual 
15-2  3700S-901-01 Rev. C / July 2008 
Environmental conditions 
Conduct the verification procedures in a location that has: 
  An ambient temperature of 18°C to 28°C (65°F to 82°F) 
  A relative humidity of less than 80%, unless otherwise noted 
Warmup period 
NOTE  At the factory, units are calibrated without any switch cards installed 
and all slots are covered with blank slot covers. The slot covers come 
installed on the unit when it is shipped. 
  If it is more convenient to calibrate the unit with switch cards installed, 
make sure all channels are open and any empty slots are covered 
with blank slot covers. 
Allow the System Switch/Multimeter to warm up for at least two hours before 
performing calibration. 
If the instrument has been subjected to temperature extremes (those outside the 
ranges stated in Environmental conditions (on page 14-2)), allow extra time for 
the instrument‟s internal temperature to stabilize. Typically, you need to allow 
one extra hour to stabilize a unit that is 10°C (18°F) outside the specified 
temperature range. 
Also, allow the test equipment to warm up for the minimum time specified by the 
manufacturer. 
Line power 
The Series 3700 requires a line voltage of 100V to 240V (±10%), and a line 
frequency of 50Hz or 60Hz. 
NOTE  The instrument automatically senses the line frequency at power-up. 