User manual
In this section: 
Overview ......................................................................... 15-1 
Environmental conditions ................................................ 15-2 
Calibration considerations ............................................... 15-3 
Calibration ....................................................................... 15-4 
Remote calibration procedure ......................................... 15-5 
Overview 
Use the procedures in this section to calibrate the Keithley Instruments Series 
3700 System Switch/Multimeter. 
WARNING  The information in this section is intended for qualified 
service personnel only. Do not attempt these procedures 
unless you are qualified to do so. 
  Some of these procedures may expose you to hazardous 
voltages, that if contacted, could cause personal injury or 
death. Use appropriate safety precautions when working 
with hazardous voltages. 
  For the plug-in modules, the maximum common-mode 
voltage (voltage between any plug-in module terminal and 
chassis ground) is 300V DC or 300V RMS. Exceeding this 
value may cause a breakdown in insulation, creating a 
shock hazard. 
All procedures in this section require accurate calibration equipment to supply 
precise DC and AC voltages, DC and AC currents, and resistance values. 
Calibration can be performed any time by an operator using the Instrument 
Control Language (ICL) commands sent either over the IEEE-488 bus or 
Ethernet. DC-only or AC-only calibration may be performed individually, if 
desired. 
Section 15 
Calibration 