User manual
Series 3700 System Switch/Multimeter Reference Manual  Section 13: Instrument Control Library (ICL) 
3700S-901-01 Rev. C / July 2008  13-303 
tspnet.idn() 
Function 
Retrieves response of remote device to '*IDN?' 
Usage 
<idn string> = tspnet.idn(<connection id>) 
idn_string: Response as a string type 
connection id: Integer value used as a handle for other tspnet commands 
Remarks 
Sends the '*idn?' string to the remote device and retrieves its response. 
Errors: 
  Invalid Specified Connection 
  Connection Not Available 
  Connection Failed, Aborted 
  Write Failed, Timeout 
  Write Failed 
  Read Failed, Timeout 
  Read Failed 
  Read Failed, Aborted 
Example 
Retrieve and print response of 'IDN?*' from the remote device: 
print(tspnet.idn(mydevice)) 
KEITHLEY INSTRUMENTS INC.,MODEL 3706,34345656,01.02a 
tspnet.read() 
Function 
Reads data from remote device. 
Usage 
[variable =] tspnet.read(<connection id>, [<format 
string>]) 
connection id: Integer value used as a handle for other tspnet commands 
format string: Definition of format string for the input field using zeros (0), the decimal 
point (.), the polarity sign (+), and 'E' for exponent. 