User manual
Section 13: Instrument Control Library (ICL)  Series 3700 System Switch/Multimeter Reference Manual 
13-298  3700S-901-01 Rev. C / July 2008 
tsplink.trigger[N].mode 
Also see 
tsplink.trigger[N].assert (on page 3-12) 
tsplink.trigger[N].clear() (on page 3-13) 
tsplink.trigger[N].overrun (on page 3-15) 
tsplink.trigger[N].release() (on page 3-15) 
tsplink.trigger[N].stimulus (on page 13-298) 
tsplink.trigger[N].wait() (on page 3-15) 
tsplink.trigger[N].overrun 
Attribute 
Event detector overrun status. 
Usage 
overrun = tsplink.trigger[N].overrun 
overrun: Trigger overrun state. 
N: The trigger line (1–3). 
Remarks 
This attribute is a read-only attribute. It indicates if an event was ignored due to the 
event detector being in the detected state when the event occurred. This is an 
indication of the state of the event detector built into the synchronization line itself. 
This attribute does not indicate if an overrun occurred in any other part of the trigger 
model or in any other construct that is monitoring the event. 
Also see 
tsplink.trigger[N].assert (on page 3-12) 
tsplink.trigger[N].clear() (on page 3-13) 
tsplink.trigger[N].mode (on page 3-13) 
tsplink.trigger[N].release() (on page 3-15) 
tsplink.trigger[N].stimulus (on page 13-298) 
tsplink.trigger[N].wait() (on page 3-15) 
tsplink.trigger[N].release() 
Function 
Release a latched trigger. 
Usage 
tsplink.trigger[N].release() 
N: The trigger line (1–3). 
Remarks 
This function will release a latched trigger on the given TSP-Link
TM
 trigger line. 
Also see 
tsplink.trigger[N].assert (on page 3-12) 
tsplink.trigger[N].clear() (on page 3-13) 
tsplink.trigger[N].mode (on page 3-13) 
tsplink.trigger[N].overrun (on page 3-15) 
tsplink.trigger[N].stimulus (on page 13-298) 
tsplink.trigger[N].wait() (on page 3-15) 