Series 3700 System Switch/Multimeter Reference Manual  Section 13: Instrument Control Library (ICL) 
3700S-901-01 Rev. C / July 2008  13-247 
scan.trigger.channel.stimulus 
Example 
To set trigger stimulus of the channel event detector to scan start event: 
scan.trigger.channel.stimulus = 
scan.trigger.EVENT_SCAN_START 
scan.trigger.clear() 
Function 
Clear the trigger model. 
Usage 
scan.trigger.clear() 
Remarks 
This function will set the channel, measure and sequence event detectors of the 
trigger model to the undetected state. 
Example 
To clear the trigger model: 
scan.trigger.clear() 
scan.trigger.measure.clear() 
Function 
Clear the measure event detector. 
Usage 
scan.trigger.measure.clear() 
Remarks 
This function will set the measure event detector of the trigger model to the undetected 
state. 
Example 
To clear the measure event detector: 
scan.trigger.measure.clear() 
scan.trigger.measure.set() 
Function 
Set the measure event detector to the detected state. 
Usage 
scan.trigger.measure.set() 
Remarks 
This function will set the measure event detector of the trigger model to the detected 
state. 
Example 
To set the measure event detector to the detected state: 
scan.trigger.measure.set() 
scan.trigger.measure.stimulus 
Attribute 
Measure event detector trigger stimulus selection. 
Usage 
To read the trigger stimulus: 
eventid = scan.trigger.measure.stimulus 
eventid: the present trigger stimulus being used for the measure event 
To write the trigger stimulus: 
scan.trigger.measure.stimulus = eventid 
eventid: the trigger stimulus to set for the measure event 