User manual
Series 3700 System Switch/Multimeter Reference Manual  Section 13: Instrument Control Library (ICL) 
3700S-901-01 Rev. C / July 2008  13-91 
digio.trigger[N].release() 
Example 
Releases trigger line 4: 
digio.trigger[4].release() 
digio.trigger[N].stimulus 
Attribute 
Event to cause this trigger to assert. 
Usage 
triggerstimulus = digio.trigger[N].stimulus 
- or - 
digio.trigger[N].stimulus = triggerstimulus 
N: Digital I/O trigger line: 1 to 14 
triggerstimulus: The event identifier for the triggering event. 
Remarks 
This attribute selects which event(s) will cause digital output line to assert a trigger. 
Set this attribute to 0 to bypass waiting for an event. 
Trigger stimulus for a digio line may be set to one of the following (existing trigger 
event IDs): 
  digio.trigger[N].EVENT_ID: An edge (either rising, falling, or either based on the 
configuration of the line) on the digital input line. 
  display.trigger.EVENT_ID: The trigger key on the front panel is pressed. 
  trigger.EVENT_ID: A *trg message on the active command interface. If GPIB is the 
active command interface, a GET message will also generate this event. 
  trigger.blender[N].EVENT_ID: A combination of events has occurred. 
  trigger.timer[N].EVENT_ID: A delay expired. 
  tsplink.trigger[N].EVENT_ID: An edge (either rising, falling, or either based on the 
configuration of the line) on the tsplink trigger line. 
  lan.trigger[N].EVENT_ID 
  scan.trigger.EVENT_SCAN_READY: Scan Ready Event. 
  scan.trigger.EVENT_SCAN_START: Scan Start Event 
  scan.trigger.EVENT_CHANNEL_READY: Channel Ready Event 
  scan.trigger.EVENT_MEASURE_COMP: Measure Complete Event 
  scan.trigger.EVENT_SEQUENCE_COMP: Sequence Complete Event 
  scan.trigger.EVENT_SCAN_COMP: Scan Complete Event 
  scan.trigger.EVENT_IDLE: Idle Event 
NOTE  Use ICL define to set the stimulus value rather than the define value. Doing 
this will make the code compatible for future upgrades because they may 
need to change when enhancements are added to the instrument. 
Example 
To set the trigger stimulus of digital I/O line 3 to be the channel ready event during a 
scan: 
digio.trigger[3].stimulus = scan.EVENT_CHANNEL_READY 
To clear the trigger stimulus of digital I/O line 3: 
digio.trigger[3].stimulus = 0 