User manual
Section 12: Status Model  Series 3700 System Switch/Multimeter Reference Manual 
12-18  3700S-901-01 Rev. C / July 2008 
Commands 
Bit 
To read registers: 
print(status.system4.enable) 
print(status.system4.condition) 
print(status.system4.event) 
To set register bits: 
status.system5.enable = status.system.EXTENSION_BIT 
B0 
status.system5.enable = status.system.EXT 
B0 
status.system5.enable = status.system.NODEn 
Bn 
To read registers: 
print(status.system5.enable) 
print(status.system5.condition) 
print(status.system5.event) 
Refer to the following table for available n values: 
Command 
n value 
status.system 
1 to 14 
status.system2 
15 to 28 
status.system3 
29 to 42 
status.system4 
43 to 56 
status.system5 
57 to 64 