User manual
Series 3700 System Switch/Multimeter Reference Manual  Section 10: TSP-Net 
3700S-901-01 Rev. C / July 2008  10-11 
tspnet.reset() 
Remarks 
This command disconnects the all devices currently connected. 
For Keithley Instruments TSP
TM
 devices, this results in any remotely running 
commands or scripts being terminated. 
Errors: 
  <none> 
Example 
tspnet.reset() 
tspnet.termination() 
Function 
Device line termination. 
Usage 
<termination type> = tspnet.termination(<connection id>, 
[<termination type>]) 
connection id: Integer value used as a handle for other tspnet commands 
termination type: tspnet.TERM_LF, tspnet.TERM_CR, 
tspnet.TERM_CRLF, or tspnet.TERM_LFCR 
Remarks 
This setting sets and gets the termination characters used to determine the end of a 
line for lines being received by a connection. It also is used to terminate lines being 
sent to a connection. Pass the optional set value to set the termination. The current 
value is always returned. There are four possible values: LF, CR, CRLF, or LFCR. For 
TSP
TM
 devices, the default is LF. For non-TSP devices, the default is CRLF. The 
termination character resets to default when a connection is terminated. 
Errors: 
  Invalid Specified Connection 
  Invalid Termination 
Example 
Set termination character: 
tspnet.termination(mydevice, tspnet.TERM_LF) 
Gets termination character and evaluates if set to LF. Response of "1" means true, set 
to <termination type>. Response of "0" means false, not set to 
<termination type>: 
print(tspnet.termination(mydevice) == tspnet.TERM_LF) 
Output: 
1.0000000e+000 
tspnet.timeout 
Attribute 
Sets timeout value for tspnet.connect(), tspnet.execute(), and 
tspnet.read() commands. 
Usage 
tspnet.timeout [= <seconds value>] 
seconds value: Value in seconds 










