User manual
Section 8: Scanning  Series 3700 System Switch/Multimeter Reference Manual 
8-24  3700S-901-01 Rev. C / July 2008 
Input characteristics: 
All rising or falling edges generate an input trigger event 
Figure 8-9: Either edge output trigger 
Output characteristics: 
The trigger.assert command generates a low pulse that is similar to the 
falling edge trigger mode. 
Understanding synchronous triggering modes 
Use the synchronous triggering modes to implement bidirectional triggering, to 
wait for one node, or to wait for a collection of nodes to complete all triggered 
actions. 
All non-Keithley instrumentation must have a trigger mode that functions similar 
to the SynchronousA or SynchronousM trigger modes. 
To use synchronous triggering, configure the triggering master to the 
SynchronousM trigger mode or the non-Keithley equivalent. Configure all other 
nodes in the test system to SynchronousA trigger mode or a non-Keithley 
equivalent. 