User manual
Series 3700 System Switch/Multimeter Reference Manual Section 3: TSP Advanced Features
3700S-901-01 Rev. C / July 2008 3-15
tsplink.trigger[N].mode
Also see
tsplink.trigger[N].assert (on page 3-12)
tsplink.trigger[N].clear() (on page 3-13)
tsplink.trigger[N].overrun (on page 3-15)
tsplink.trigger[N].release() (on page 3-15)
tsplink.trigger[N].stimulus (on page 13-298)
tsplink.trigger[N].wait() (on page 3-15)
tsplink.trigger[N].overrun
Attribute
Event detector overrun status.
Usage
overrun = tsplink.trigger[N].overrun
overrun: Trigger overrun state.
N: The trigger line (1–3).
Remarks
This attribute is a read-only attribute. It indicates if an event was ignored due to the
event detector being in the detected state when the event occurred. This is an
indication of the state of the event detector built into the synchronization line itself.
This attribute does not indicate if an overrun occurred in any other part of the trigger
model or in any other construct that is monitoring the event.
Also see
tsplink.trigger[N].assert (on page 3-12)
tsplink.trigger[N].clear() (on page 3-13)
tsplink.trigger[N].mode (on page 3-13)
tsplink.trigger[N].release() (on page 3-15)
tsplink.trigger[N].stimulus (on page 13-298)
tsplink.trigger[N].wait() (on page 3-15)
tsplink.trigger[N].release()
Function
Release a latched trigger.
Usage
tsplink.trigger[N].release()
N: The trigger line (1–3).
Remarks
This function will release a latched trigger on the given TSP-Link
TM
trigger line.
Also see
tsplink.trigger[N].assert (on page 3-12)
tsplink.trigger[N].clear() (on page 3-13)
tsplink.trigger[N].mode (on page 3-13)
tsplink.trigger[N].overrun (on page 3-15)
tsplink.trigger[N].stimulus (on page 13-298)
tsplink.trigger[N].wait() (on page 3-15)