Series 3700 System Switch/Multimeter Reference Manual Section 15: Calibration
3700S-901-01 Rev. C / July 2008 15-13
AC Cal Step 10: 300V @ 1kHz
1. Send the following command:
SEND dmm.range = 300
2. Source 300V @ 1kHz
3. Send the following command:
SEND dmm.calibration.ac(10) print("done")
AC current calibration
Make the connections as shown, then perform the following calibration steps
(AC Cal Step 11 through Step 16):
Figure 15-6: AC current calibration 1mA to 1A range
AC Cal Step 11: 100µA @ 1kHz
1. Send the following commands:
SEND dmm.func = dmm.ac_current
SEND dmm.range = 100e-6
2. Source 100µA @ 1kHz.
3. Send the following command:
SEND dmm.calibration.ac(11) print("done")
AC Cal Step 12: 1mA @ 1kHz
1. Send the following command:
SEND dmm.range = 1e-3
2. Source 1mA @ 1kHz.
3. Send the following command:
SEND dmm.calibration.ac(12) print("done")