Series 3700 System Switch/Multimeter Reference Manual Section 15: Calibration
3700S-901-01 Rev. C / July 2008 15-9
DC Cal Step 8: 100k Ohm
1. Send the following command:
SEND dmm.range = 100e+3
2. Source 100k Ohm, and then read the resistor value from the calibrator.
3. Send the following command:
SEND dmm.calibration.dc(8, (resistor value))
print("done")
DC Cal Step 9: 1M Ohm
1. Send the following command:
SEND dmm.range = 1e+6
2. Source 1M Ohm then read the resistor value from the calibrator.
3. Send the following command:
SEND dmm.calibration.dc(9, (resistor value))
print("done")
DC current calibration
Make the connections as shown, then perform the following calibration steps
(DC Cal Step 10 through Step 14):
Figure 15-4: DC current calibration