User manual

Section 14: Verification Series 3700 System Switch/Multimeter Reference Manual
14-20 3700S-901-01 Rev. C / July 2008
4-wire resistance verification data
Use the following values to verify the performance of the Series 3700. Actual
values depend on published specifications (see Calculating resistance reading
limits (on page 14-4)).
Connect to the Fluke 5700A calibrator
Description
Range (Ohms)
Test point
(Ohms)
Lower limit (Ohms)
Upper limit
(Ohms)
Verify 4W Res 100 Ohm
*
1.00E+02
1.00E+02
9.999310E+01
1.000069E+02
Verify 4W Res 1k Ohm
1.00E+03
1.00E+03
9.999360E+02
1.000064E+03
Verify 4W Res 10k Ohm
1.00E+04
1.00E+04
9.999360E+03
1.000064E+04
Verify 4W Res 100k Ohm
1.00E+05
1.00E+05
9.999360E+04
1.000064E+05
Verify 4W Res 1M Ohm
1.00E+06
1.00E+06
9.999560E+05
1.000044E+06
Verify 4W Res 10M Ohm
1.00E+07
1.00E+07
9.995900E+06
1.000410E+07
NOTE The asterisk (*) designates the ranges that offset compensation is
being used.