User manual
Section 14: Verification Series 3700 System Switch/Multimeter Reference Manual
14-18 3700S-901-01 Rev. C / July 2008
Verifying frequency
To verify the Series 3700 frequency function:
1. Connect the Agilent 33220A function generator to the Series 3700 INPUT
pins.
2. Set the function generator to output a 1kHz, 5V RMS sine wave.
3. Select the Series 3700 frequency function by pressing the FREQ key.
4. Verify that each Series 3700 frequency reading is within the limits contained
in the table contained in Frequency verification data (on page 14-18).
Figure 14-8: Frequency verification
Frequency verification data
Use the following values to verify the performance of the Series 3700. Actual
values depend on published specifications (see Example reading limit
calculation (on page 14-4)).
Connect the Agilent 33220A Generator
Description
Range
(V)
Frequency (Hz)
Lower limit (Hz)
Upper limit
(Hz)
Verify Frequency 1kHz
1.00E+01
1.00E+03
9.999167E+02
1.000083E+03
Verify Frequency 10kHz
1.00E+01
1.00E+04
9.999167E+03
1.000083E+04
Verify Frequency
100kHz
1.00E+01
1.00E+05
9.999167E+04
1.000083E+05
Verify Frequency
250kHz
1.00E+01
2.50E+05
2.499797E+05
2.500203E+05
Verify Frequency
500kHz
1.00E+01
5.00E+05
4.999597E+05
5.000403E+05