Series 3700 System Switch/Multimeter Reference Manual Section 14: Verification
3700S-901-01 Rev. C / July 2008 14-17
Connect to the Fluke 5700A calibrator
Description
Range
(A)
Test point
(A)
Lower limit (A)
Upper limit
(A)
Verify AC Curr 1mA @ 1kHz
1.00E-03
1.00E-03
9.989000E-04
1.001100E-03
Verify AC Curr 1mA @ 5kHz
1.00E-03
1.00E-03
9.989000E-04
1.001100E-03
Verify AC Curr 10mA @ 40Hz
1.00E-02
1.00E-02
9.989000E-03
1.001100E-02
Verify AC Curr 10mA @ 1kHz
1.00E-02
1.00E-02
9.989000E-03
1.001100E-02
Verify AC Curr 10mA @ 5kHz
1.00E-02
1.00E-02
9.989000E-03
1.001100E-02
Verify AC Curr 100mA @
40Hz
1.00E-01
1.00E-01
9.989000E-02
1.001100E-01
Verify AC Curr 100mA @
1kHz
1.00E-01
1.00E-01
9.989000E-02
1.001100E-01
Verify AC Curr 100mA @
5kHz
1.00E-01
1.00E-01
9.989000E-02
1.001100E-01
Verify AC Curr 1A @ 40Hz
1.00E+00
1.00E+00
9.977000E-01
1.002300E+00
Verify AC Curr 1A @ 1kHz
1.00E+00
1.00E+00
9.977000E-01
1.002300E+00
Verify AC Curr 1A @ 5kHz
1.00E+00
1.00E+00
9.977000E-01
1.002300E+00
AC current verification data 3A range
Use the following values to verify the performance of the Series 3700. Actual
values depend on published specifications (see Example reading limit
calculation (on page 14-4)).
Connect to the Fluke 5725A amplifier
Description
Range
(A)
Test point
(A)
Lower limit (A)
Upper limit
(A)
Verify AC Curr 3A @
40Hz
3.00E+00
3.00E+00
2.993100E+00
3.006900E+00
Verify AC Curr 3A @
1kHz
3.00E+00
3.00E+00
2.993100E+00
3.006900E+00
Verify AC Curr 3A @
5kHz
3.00E+00
3.00E+00
2.993100E+00
3.006900E+00