User manual
Section 14: Verification Series 3700 System Switch/Multimeter Reference Manual
14-14 3700S-901-01 Rev. C / July 2008
Figure 14-5: DC current verification 3A range diagram
DC current verification data 1mA to 3A ranges
Use the following values to verify the performance of the Series 3700. Actual
values depend on published specifications (see Example reading limit
calculation (on page 14-4)).
Remove HP3458A, only connect the 5700
Description
Range
(A)
Test point
(A)
Lower limit (A)
Upper limit
(A)
Verify 1mA Zero
1.00E-03
0.00E+00
-9.000000E-09
9.000000E-09
Verify DC Curr 1mA
1.00E-03
1.00E-03
9.994910E-04
1.000509E-03
Verify DC Curr 1mA
1.00E-03
-1.00E-03
-1.000509E-03
-9.994910E-04
Verify 10mA Zero
1.00E-02
0.00E+00
-9.000000E-08
9.000000E-08
Verify DC Curr 10mA
1.00E-02
1.00E-02
9.994910E-03
1.000509E-02
Verify DC Curr 10mA
1.00E-02
-1.00E-02
-1.000509E-02
-9.994910E-03
Verify 100mA Zero
1.00E-01
0.00E+00
-9.000000E-07
9.000000E-07
Verify DC Curr
100mA
1.00E-01
1.00E-01
9.994910E-02
1.000509E-01
Verify DC Curr
100mA
1.00E-01
-1.00E-01
-1.000509E-01
-9.994910E-02
Verify DC Curr 1A
1.00E+00
1.00E+00
9.991900E-01
1.000810E+00