Section 14: Verification Series 3700 System Switch/Multimeter Reference Manual
14-12 3700S-901-01 Rev. C / July 2008
5. Set the calibrator to source zero current and rel both the Series 3700 and
the HP3458A.
6. Source DC current for each of the test points summarized in the DC voltage
verification data (on page 14-7) table. For each setting, be sure that the
reading is within stated limits.
Figure 14-3: DC current verification 10µA to 100µA ranges
DC current verification data 10µA to 100µA ranges
Use the following values to verify the performance of the Series 3700. Actual
values depend on published specifications (see Example reading limit
calculation (on page 14-4)).
Connect HP3458A in series with 5700 calibrator
Description
Range (A)
Test
point(A)
Lower limit (A)
Upper limit (A)
Verify 10µA Zero
1.00E-05
0.00E+00
-3.000000E-10
3.000000E-10
Verify DC Curr 10µA
1.00E-05
1.00E-05
9.994700E-06
1.000530E-05
Verify DC Curr 10µA
1.00E-05
-1.00E-05
-1.000530E-05
-9.994700E-06