Section 14: Verification Series 3700 System Switch/Multimeter Reference Manual
14-10 3700S-901-01 Rev. C / July 2008
Figure 14-2: AC voltage verification
ACV verification data
Use the following values to verify the performance of the Series 3700. Actual
values depend on published specifications (see Example reading limit
calculation (on page 14-4)).
Connect to the Fluke 5700A calibrator
Description
Range (V)
Test point
(V)
Lower limit (V)
Upper limit (V)
Verify ACV 100mV @ 20Hz
1.00E-01
1.00E-01
9.897000E-02
1.010300E-01
Verify ACV 100mV @ 1kHz
1.00E-01
1.00E-01
9.992000E-02
1.000800E-01
Verify ACV 100mV @ 50kHz
1.00E-01
1.00E-01
9.984000E-02
1.001600E-01
Verify ACV 100mV @
100kHz
1.00E-01
1.00E-01
9.932000E-02
1.006800E-01
Verify ACV 1V @ 20Hz
1.00E+00
1.00E+00
9.992000E-01
1.000800E+00
Verify ACV 1V @ 1kHz
1.00E+00
1.00E+00
9.992000E-01
1.000800E+00
Verify ACV 1V @ 50kHz
1.00E+00
1.00E+00
9.984000E-01
1.001600E+00
Verify ACV 1V @ 100kHz
1.00E+00
1.00E+00
9.932000E-01
1.006800E+00
Verify ACV 10V @ 1kHz
1.00E+01
1.00E+01
9.992000E+00
1.000800E+01
Verify ACV 10V @ 50kHz
1.00E+01
1.00E+01
9.984000E+00
1.001600E+01