User manual

Section 14: Verification Series 3700 System Switch/Multimeter Reference Manual
14-8 3700S-901-01 Rev. C / July 2008
Connect to the Fluke 5700A Calibrator
Description
Range
(V)
Test point
(V)
Lower limit (V)
Upper limit (V)
Verify DCV
100mV
1.00E-01
5.00E-02
4.999760E-02
5.000240E-02
Verify DCV
100mV
1.00E-01
-5.00E-02
-5.000240E-02
-4.999760E-02
Verify DCV
100mV
1.00E-01
-1.00E-01
-1.000039E-01
-9.999610E-02
Rel Series 3700
1.00E+00
0.00E+00
N/A
N/A
Verify DCV 1V
1.00E+00
1.00E+00
9.999680E-01
1.000032E+00
Verify DCV 1V
1.00E+00
5.00E-01
4.999830E-01
5.000170E-01
Verify DCV 1V
1.00E+00
-5.00E-01
-5.000170E-01
-4.999830E-01
Verify DCV 1V
1.00E+00
-1.00E+00
-1.000032E+00
-9.999680E-01
Verify DCV 10V
1.00E+01
1.00E+01
9.999730E+00
1.000027E+01
Verify DCV 10V
1.00E+01
5.00E+00
4.999855E+00
5.000145E+00
Verify DCV 10V
1.00E+01
0.00E+00
-2.000000E-05
2.000000E-05
Verify DCV 10V
1.00E+01
-5.00E+00
-5.000145E+00
-
4.999855E+00
Verify DCV 10V
1.00E+01
-1.00E+01
-1.000027E+01
-
9.999730E+00
Verify DCV 100V
1.00E+02
1.00E+02
9.999540E+01
1.000046E+02
Verify DCV 100V
1.00E+02
5.00E+01
4.999740E+01
5.000260E+01
Verify DCV 100V
1.00E+02
0.00E+00
-6.000000E-04
6.000000E-04
Verify DCV 100V
1.00E+02
-5.00E+01
-5.000260E+01
-
4.999740E+01
Verify DCV 100V
1.00E+02
-1.00E+02
-1.000046E+02
-
9.999540E+01