User manual

Section 13: Instrument Control Library (ICL) Series 3700 System Switch/Multimeter Reference Manual
13-270 3700S-901-01 Rev. C / July 2008
status.node_event
Example
Reads the status node event register:
nodeeventreg = status. node_event
print(nodeeventreg )
Output: 1.29000e+02
The above output indicates that bits B0 (MSB) and B7 (OSB) are set.
status.operation.*
.condition
.enable
.event
.ntr
.ptr
Attribute
Operation event status register set.
Usage
To read condition, enable, event, NTR and PTR registers:
operreg = status.operation.condition
operreg = status.operation.enable
operreg = status.operation.event
operreg = status.operation.ntr
operreg = status.operation.ptr
To write to enable, NTR and PTR registers:
status.operation.enable = operreg
status.operation.ntr = operreg
status.operation.ptr = operreg
Set operreg to one of the following values:
0 clears all bits.
To set CAL bit (B0):
status.operation.CALIBRATING
or
status.operation.CAL
To set MEAS bit (B4):
status.operation.MEASURING
or
status.operation.MEAS
To set PRMPTS bit (B11):
status.operation.PROMPTS
- or -
status.operation.PRMPTS
To set USER bit (B12): status.operation.USER
To set PROG bit (B14):
status.operation.PROGRAM_RUNNING
- or -
status.operation.PROG