User manual

Series 3700 System Switch/Multimeter Reference Manual Section 13: Instrument Control Library (ICL)
3700S-901-01 Rev. C / July 2008 13-247
scan.trigger.channel.stimulus
Example
To set trigger stimulus of the channel event detector to scan start event:
scan.trigger.channel.stimulus =
scan.trigger.EVENT_SCAN_START
scan.trigger.clear()
Function
Clear the trigger model.
Usage
scan.trigger.clear()
Remarks
This function will set the channel, measure and sequence event detectors of the
trigger model to the undetected state.
Example
To clear the trigger model:
scan.trigger.clear()
scan.trigger.measure.clear()
Function
Clear the measure event detector.
Usage
scan.trigger.measure.clear()
Remarks
This function will set the measure event detector of the trigger model to the undetected
state.
Example
To clear the measure event detector:
scan.trigger.measure.clear()
scan.trigger.measure.set()
Function
Set the measure event detector to the detected state.
Usage
scan.trigger.measure.set()
Remarks
This function will set the measure event detector of the trigger model to the detected
state.
Example
To set the measure event detector to the detected state:
scan.trigger.measure.set()
scan.trigger.measure.stimulus
Attribute
Measure event detector trigger stimulus selection.
Usage
To read the trigger stimulus:
eventid = scan.trigger.measure.stimulus
eventid: the present trigger stimulus being used for the measure event
To write the trigger stimulus:
scan.trigger.measure.stimulus = eventid
eventid: the trigger stimulus to set for the measure event