Section 13: Instrument Control Library (ICL) Series 3700 System Switch/Multimeter Reference Manual
13-140 3700S-901-01 Rev. C / July 2008
dmm.getconfig()
Remarks
Use this command to query for DMM configurations associated with channels or
channel patterns. The response will be a comma-delimited string listing the
configurations in the same order as specified in ch_list.
The configurations listed in the response indicate how the DMM will be configured
when the corresponding channel or channel pattern is closed with the dmm.close() (on
page 13-123) function or used in a scan list without an overriding DMM configuration.
An error will be generated if:
Syntax error exists in parameter string.
An empty parameter string is specified.
A specified channel or channel pattern is invalid.
Channel number does not exist for slot based on installed card.
Slot is empty.
Channel pattern does not exist.
Channel being specified does not support a configuration setting like a digital I/O
channel or analog backplane relay.
Command processing will stop once an error is detected and a nil response is
generated.
Also see
dmm.setconfig() (on page 13-168)
Example
To query channels on Slots 1 and 2:
myconfigs = dmm.getconfig('slot1, slot2')
To query Channels 1 to 10 on Slot 3:
myconfigs = dmm.getconfig('3001:3010')
dmm.inputdivider
Attribute
Enables or disables the 10M ohm input divider.
Usage
To read the input divider state:
value = dmm.inputdivider
value: Represents the present input divider
To write the input divider state:
dmm.inputdivider = value
value: Represents the desired input divider state.
Use one of the following for value:
dmm.ON or 1
dmm.OFF or 0