User manual
Series 3700 System Switch/Multimeter Reference Manual Section 8: Scanning
3700S-901-01 Rev. C / July 2008 8-21
Output characteristics:
The trigger.assert command generates a low pulse for the programmed
pulse duration.
Rising edge master trigger mode (version 1.4.0 or higher)
Use the rising edge master trigger mode (RisingM) to synchronize with
non-Keithley Instruments that require a high pulse. Input trigger detection is not
available in this trigger mode. You can use the RisingM trigger mode to generate
rising edge pulses.
NOTE The RisingM trigger mode does not function properly if the line is
driven low by an external drive.
Figure 8-5: RisingM output trigger
Output characteristics:
The trigger.assert command causes the physical line state to float high
during the trigger pulse duration.