User manual

Section 7: Buffer: Data Storage and Retrieval Series 3700 System Switch/Multimeter Reference Manual
7-4 3700S-901-01 Rev. C / July 2008
To save readings to a USB flash drive:
1. Select a non-empty reading buffer.
2. Press the STORE key. The [BUFFER NAME] MENU is displayed.
3. Select the SAVE menu item, and press the ENTER key. The SAVE RD
BUFFER menu is displayed.
4. Press the ENTER key when USB is highlighted.
5. Using the navigation wheel and CURSOR keys, enter the filename
where the data will be saved on the installed USB flash drive. The starting
name is:
<reading buffer name>_nn_ _ _
Where: nn starts at 01 and automatically increments. For example, if the
selected reading buffer is fpbuf1, then the starting name is fpbuf1_01_
_ _.
6. Press the ENTER key to save the data to the installed USB flash drive or the
EXIT key to cancel.
Clearing readings
When clearing readings, you must select a reading buffer. See Selecting a
reading buffer (on page 7-3) for more information.
To clear readings:
1. Select a reading buffer.
2. Press the STORE key. The [BUFFER NAME] MENU is displayed.
3. Select the CLEAR menu item, and press the ENTER key.
4. At the prompt, select YES or NO and press the ENTER key.