Section 5: Range, Digits, Rate, Bandwidth, and Filter Series 3700 System Switch/Multimeter Reference Manual
5-2 3700S-901-01 Rev. C / July 2008
Function
Ranges
Maximum reading
TMP (temperature)
-200°C to 1820°C
Sensor dependent
FREQ (frequency)
100mV, 1V, 10V, 100V, 300V
3Hz to 500kHz
PER (period)
100mV, 1V, 10V, 100V, 300V
2µs to 333ms
CNT (continuity)
1k
Threshold adjustable 1 to
1000
CSΩ (common-side ohm)
1 , 10 , 100 , 1k , 10k ,
100k , 1M , 10M , 100M
120M
Temperature
There is no range selection for temperature measurements, which are
performed on a single fixed range. Depending on the sensor, the maximum
temperature readings range from -200°C to 1820°C.
Manual range keys
To change range, press the RANGE or ▼ key. The instrument changes one
range value of the active function per key press. The selected range is displayed
in the attribute list on the second line of the front panel display.
NOTE The manual range keys have no effect on temperature
measurements.
If the instrument displays the "Overflow" message on a particular range, select a
higher range until an on-range reading is displayed. For best accuracy and
resolution, use the lowest range available that does not cause an overflow.