User manual

In this section:
Range ............................................................................... 5-1
Digits ICL programming .................................................... 5-4
Rate .................................................................................. 5-5
Bandwidth ......................................................................... 5-7
Filter .................................................................................. 5-8
Range
The range setting is "remembered" by each measurement function. Selecting a
function returns the instrument to the last range setting for that function. You
cannot select a range that includes different channel types.
NOTE A power cycle or an instrument will clear "remembered" ranges.
Measurement ranges and maximum readings
The range that is selected affects both measurement accuracy as well as the
maximum measurable level. Input values that exceed the maximum readings
cause an "Overflow" message to be displayed.
Function
Ranges
Maximum reading
DCV (DC voltage)
100mV, 1V, 10V, 100V, 300V
± 303V
ACV (AC voltage)
100mV, 1V, 10V, 100V, 300V
± 303V
DCI (DC current)
10µA, 100µA, 1mA, 10mA,
100mA, 1A
± 3.1A
ACI (AC current)
1mA, 10mA, 100mA, 1A
± 3.1A
2 (2-wire ohm)
10 , 100 , 1k , 10k ,
100k , 1M , 10M , 100M
120M
4 (4-wire ohm)
1 , 10 , 100 , 1k , 10k ,
100k , 1M , 10M , 100M
120M
4 OC (4-wire ohm offset
compensated)
1 , 10 , 100 , 1k , 10k
12k
4 DRY+ (4-wire ohm dry
circuit)
1 , 10 , 100 , 1k , 10k
2.4k
Section 5
Range, Digits, Rate, Bandwidth, and Filter