Series 3700 System Switch/Multimeter Reference Manual Section 4: Using the Front Panel
3700S-901-01 Rev. C / July 2008 4-25
UNITS
Configures the units for voltage and temperature measurements. Related ICL
command: dmm.units (on page 13-174).
ENTER key
Press the ENTER key to accept the current selection or bring up the next menu
options.
NOTE Pressing the navigation wheel performs the same function as the
ENTER key.
EXIT key
Press the EXIT key to:
Cancel the selection and to return to the previous menu display.
Exit remote operation.
Abort a scan that is running.
Abort a script that is executing.
FILTER key
Press the FILTER key to enable and disable the filter for selected function.
When the filter is enabled, the FILT annunciator will light. See Filter (on page 5-
8) for more information.
FILTER key configuration
Press the CONFIG key and then the FILTER key to open the FILTER menu.
The FILTER menu contains the following menu items:
ENABLE: Enables or disables filtered measurements for the selected DMM
function. Related ICL command: dmm.filter.enable (on page 13-134).
COUNT: Indicates the filter count setting for the selected DMM function.
Related ICL command: dmm.filter.count (on page 13-134).
TYPE: Indicates the filter averaging type for the DMM measurements on the
selected DMM functions (MOVING or REPEAT). Related ICL command:
dmm.filter.type (on page 13-135).
WINDOW: Indicates the filter window for the DMM measurements (0 to 10%
in 0.1% increments). Related ICL command: dmm.filter.window (on page 13-
136).