3700S-901-01 Rev. C / July 2008 4-17
POWER switch
Press this switch to turn the Series 3700 on (I). Press it again to turn the Series
3700 off (O).
RESET switch
Use the RESET switch to restore the Series 3700 factory default LAN settings.
Refer to the LAN functions and attributes (on page 13-190) (lan.config.x,
where x represents the specific command) for factory default information.
Operation keys
CHAN key
Different menus are displayed for switch channel types or non-switch channel
types when the CHAN key is pressed.
CHAN key - switch channel type
Press the CHAN key to open the CHANNEL ACTION menu.
The CHANNEL ACTION menu contains the following items:
OPEN: This menu item opens the specified channels for switching aspects.
Related Instrument Control Library (ICL) command: channel.open() (on
page 13-59).
CLOSE: This menu item closes specified channels. These closures are
appended to the already closed channels. Related ICL command:
channel.close() (on page 13-37).
EXCLOSE: This menu item closes the specified channels so that they are
exclusively closed. Related ICL command: channel.exclusiveclose() (on
page 13-41).
EXSLOTCLOSE: This menu item exclusively closes specified channels on
the specified slots. Related ICL command: channel.exclusiveslotclose() (on
page 13-43).
RESET: This menu item resets specified channels to factory default
settings. Resetting a channel deletes any channel patterns that contain that
channel. Related ICL command: channel.reset() (on page 13-68).
CHAN key - DIGIO channel type
Press the CHAN key to open the DIGIO ACTION menu. Unless noted, the menu
option supports a range of selected channels.