Datasheet

SWITCHING AND CONTROL
www.keithley.com
1.888.KEITHLEY
(U.S. only)
A Greater Measure of Confidence
AC MEASUREMENT CHARACTERISTICS
AC VOLTS
MEASUREMENT METHOD: AC-coupled, True RMS.
INPUT IMPEDANCE: 1MW ±2% // by <150pF.
INPUT PROTECTION: 300VDC or 300Vrms rear inputs or 37xx cards.
AC CURRENT
MEASUREMENT METHOD: AC-coupled, True RMS.
Range
3 A 1 A 100 mA 10 mA 1 mA
Shunt Resistance
guaranteed by design
0.05 W 0.05 W 1.0 W 10 W 100 W
Burden Voltage
Rear Panel
<1.75 V rms <0.55 V rms <0.4 V rms <150 mV rms <125 mV rms
Burden Voltage
3721 Card
<2.4 V rms <1.0 V rms <0.6 V rms <200 mV rms <130 mV rms
INPUT PROTECTION: 3A, 250V fuse.
FREQUENCY AND PERIOD
MEASUREMENT METHOD: Reciprocal Counting technique.
GATE TIME: dmm.aperture=0.2730.01. Default 0.01s.
AC GENERAL
AC CMRR
6
: 70dB.
VOLT·HERTZ PRODUCT: 8×10
7
Volt·Hz (guaranteed by design), 2.1×10
7
Volt·Hz verified. Input
frequency verified for 3×10
5
Hz.
AC NOTES
1. 20% overrange on AC functions except 1% on 300V and 3.33% on 3A. Default resolution is 5½ digits, maximum
useable resolution is 6½ with 7½ digits programmable.
2. Specification are for Detector Bandwidth 3 and sinewave inputs >5% of range. Detector Bandwidth 3 and 30
are multi-sample A/D conversions. Detector bandwidth 300 is a single A/D conversion, programmable from
0.0005PLC to 15PLC. Default condition set to 1PLC.
3. Applies to 0°–18°C and 28°–50°C.
4. Specified for square wave inputs. Input signal must be >10% of ACV range. If input is <20mV on the 100mV
range then the frequency must be >10Hz. For sinewave inputs, frequency must be >100Hz.
5. Applies to non-sinewave inputs 5Hz–>10kHz, and DC content 3% of range.
6. For 1k
W
unbalance in LO lead.
7. For Model 3721, 1mA ACI, add 0.05% to “of reading” uncertainty from 250Hz
10kHz.
8. Shunt resistance guaranteed by design.
9. Reading rates are for 60Hz (50Hz) operation using factory defaults operating conditions dmm.reset(“all”),
Autorange off, dmm.autodelay=dmm.OFF, dmm.opendetector=dmm.OFF, format.data.=format.SREAL. Ranges
as follows: DCV = 10V, 2W
W
/4W
W
= 1k
W
, DCI = 1mA, Dry-Ckt
W
= 10
W
, ACI = 1mA, and ACV = 1V. For Dry-
Ckt
W
with Offset Comp OFF 2k
W
, 60 rdg/s max. Dry-Ckt
W
with Offset Comp ON 2k
W
, 29.5 rdg/s max. For
temperature reading rates use DCV for T/C and 2W
W
for Thermistor. Speeds are typical and include measure-
ments and data transfer out the Ethernet, GPIB, or USB.
10. DMM configured for single reading, dmm.measurecount=1, and print(dmm.measure()). May require additional
settling delays for full accuracy, depending on measurement configuration.
11. DMM configured for multisample readings and single buffer transfer, dmm.measurecount=1000,
buf=dmm.makebuffer(1000), dmm.measure(buf), and printbuffer(1,1000,buf).
Series 3700A
System Switch/Multimeter
and Plug-In Cards
AC SPEEDS
Single Channel, 60Hz (50Hz) Operation
Detector
Bandwidth
Measurements into Buffer
9
(rdg/s) Measurement to PC
9
(ms/rdg)
Function NPLC Aperture (ms) Digits Azero On Azero Off Ethernet GPIB USB
ACI / ACV
3 N/A N/A 0.45 (0.45) N/A 2150 (2150) 2150 (2150) 2150 (2150)
30 N/A N/A 2.5 (2.5) N/A 400 (400) 400 (400) 400 (400)
300 1.0
10
16.67 (20) 42 (33) 59.5 (50) 19.4 (22.7) 19.5 (22.8) 19.8 (23.1)
300 0.2
10
3.33 (4.0) 120 (100) 295 (235) 7.6 (8.3) 6.2 (6.8) 6.4 (7.0)
300 0.06
11
1.0 (1.2) 170 (165) 935 (750) 1.40 (1.80) 1.50 (1.80) 1.60 (2.30)
300 0.006
11
0.100 (0.120) 218 (215) 6,200 (5,500) 0.55 (0.57) 0.65 (0.67) 0.75 (0.77)
300 0.0005
11
0.0083 (0.001) 218 (215) 14,600 (14,250) 0.50 (0.5) 0.60 (0.60) 0.70 (0.70)
Frequency/Period
N/A N/A 10–273 N/A
2× input period
+ gate time
N/A
2× input period +
gate time + 2.7ms
2× input period +
gate time + 2.8ms
2× input period +
gate time + 3.1ms
Series 3700A specifications
Series 3700A specifications