GENERAL
EXPANSION SLOTS: 6.
POWER LINE: Universal, 100V to 240V.
LINE FREQUENCY: 50Hz and 60Hz, automatically sensed at power-up.
POWER CONSUMPTION: 28VA with DMM and display, up to 140VA with six 37xx cards.
REAL TIME CLOCK: Battery backed, 10 years typical life.
EMC: Conforms to European Union EMC Directive.
SAFETY: Conforms to European Union Low Voltage Directive.
VIBRATION: MIL-PRF-28800F Class 3, Random.
WARM-UP: 2 hours to rated accuracy.
DIGITAL I/O: 25-pin female D-shell.
I/O 1–9 I/O 1014 Vext
I
SINK
, max.
5 mA 250 mA
Absolute V
IN
5.25 V to –0.25 V 5.25 V to –0.25 V 5 V to 33 V
V
IH
min
2.2 V 2.2 V
V
IL
max
0.7 V 0.7 V
V
OL
max at 5mA I
sink
0.7 V 0.7 V
V
OL
max at I
sink
max
2.3 V
V
OH
min, 0.4mA source
2.7 V 2.4 V
Min V
IN
pulse
2 µs 10 µs
Min V
O
pulse
1 µs 50 µs
5k
+5V
100
Read
Write
IN/OU
T
GND
I/O 1–9
I/O 10–14
GND
5k
+5V
100
Read
Write
I
N/OUT
Vext
TRIGGERING AND MEMORY:
Window Filter Sensitivity: 0.01%, 0.1%, 1%, 10%, or full-scale of range (none).
Trigger Delay: 0 to 99 hrs. (10µs step size).
External Trigger Delay: <10µs.
Memory: Up to 650,000 time-stamped readings with Web page disabled. Additional memory
available with external “thumb drive.”
Non-volatile Memory: Single user save setup, with up to 75 DMM configurations and 600
channel patterns (dependent on name length, DMM function and configuration, and
pattern image size). Additional memory available with external “thumb drive.
MATH FUNCTIONS: Rel, dB, Limit Test, %, 1/x, and mX+b with user defined displayed.
REMOTE INTERFACE:
Ethernet: RJ-45 connector, LXI Class B Version 2, 10/100BT, no auto MDIX.
GPIB: IEEE-488.1 compliant. Supports IEEE-488.2 common commands and status model
topology.
USB Device (rear panel, type B): Full speed, USBTMC compliant.
USB Host (front panel, type A): USB 2.0, support for thumb drives.
LXI COMPLIANCE: LXI Class B Version 2 with IEEE 1588 precision time protocol.
LXI TIMING (applies to scanning) and SPECIFICATION:
Receive LAN[0–7] Event Delay: n/s (not specified) min., 80s typ., n/s max.
Alarm to Trigger Delay: 25µs min., 50µs typ., n/s max.
Generate LAN[07] Event: n/s min., 80s typ., n/s max. (minimums are probabilistic and
represent a 95% confidence factor).
Clock Accuracy: 25ppm.
Synchronization Accuracy: <150ns (probabilistic and represents a 95% confidence factor).
Timestamp Accuracy: 100µs.
Timestamp Resolution: 20ns.
LANGUAGE: Embedded Test Script Processor (TSP) accessible from any host interface.
Responds to individual Instrument Control Library (ICL) commands. Responds to high-speed
test scripts comprised of ICL commands and Test Script Language (TSL) statements (e.g.,
branching, looping, math, etc.). Able to execute high-speed test scripts stored in memory
without host intervention.
IP CONFIGURATION: Static or DHCP.
PASSWORD PROTECTION: 11 characters
MINIMUM PC HARDWARE: Intel Pentium 3, 800MHz, 512Mbyte RAM, 210Mbyte disk space
or better.
OPERATING SYSTEMS/SOFTWARE: Windows
®
2000 and XP compatible, supports Web
browsers with Java plug-in (requires Java plug-in 1.6 or higher). Web pages served by 3706A.
OPERATING ENVIRONMENT: Specified for 0° to 50°C, 80%RH at 35°C, altitude up to
2000 meters.
STORAGE ENVIRONMENT:40° to 70°C.
DIMENSIONS:
Rack Mounted: 89mm high × 483mm wide × 457mm deep (3.5 in. × 19 in. × 18 in.).
Bench Configuration (includes handle and feet): 104mm high × 483mm wide × 457mm
deep (4.125 in. × 19 in. × 18 in.)
SHIPPING WEIGHT: 13kg (28 lbs).
Series 3700A
System Switch/Multimeter
and Plug-In Cards
Series 3700A specifications
Series 3700A specifications