User manual
3390-900-01 Rev. C / January 2009 Return to Section Topics 5-19
Model 3390 Arbitrary Waveform Generator User’s Manual Section 5: Waveform Output Operations
Figure 5-16:
Press the soft keys corresponding with frequency sweep
4. Select a start frequency and a stop frequency, or select a center frequency and a frequency
span:
• Start and stop frequency: Press the soft key under Start / Cntr to specify a start
frequency (ensure Start is highlighted, Figure 5-16). Use the navigation wheel with the
cursor keys to change the value, and then press another soft key to finish the entry and
exit. Or use the numeric keypad to enter a desired value and then press the soft key
under the desired units to finish the entry, or the soft key under Cancel to cancel the
changes. Press the soft key under Stop / Span to specify a stop frequency. Follow the
same value entry procedures.
• Center frequency and frequency span: Press the soft key under Start / Cntr to specify a
center frequency (ensure Cntr is highlighted). Use the navigation wheel with the cursor
keys to change the value, and then press another soft key to finish the entry and exit. Or
use the numeric keypad to enter a desired value and then press the soft key under the
desired units to finish the entry, or the soft key under Cancel to cancel the changes.
Press the soft key under Stop / Span to specify a span frequency. Follow the same
value entry procedures.
5. Press the soft key under Sweep / Time. Use the navigation wheel with the cursor keys to
change the value, and then press another soft key to finish the entry and exit. Or use the
numeric keypad to enter a desired value and then press the soft key under the desired units
to finish the entry, or the soft key under Cancel to cancel the changes.
6. Press the soft key under Mkr to toggle between Freq and OFF. Use the navigation wheel
with the cursor keys to enter a frequency value, and then press another soft key to finish the
entry and exit. Or use the numeric keypad to enter a desired value and then press the soft
key under the desired units to finish the entry, or the soft key under Cancel to cancel the
changes.
7. Press the soft key under Trig / Setup to configure the trigger setting (Figure 5-17). Press
the soft key under Src to select among Int (internal), Ext (external), or Man (manual) for the
trigger source.
• Internal or manual source: Press the Trig / Out soft key, then select among OFF, rising-
edge, or falling edge trigger.
• External source: Press the Slope soft key, then select rising edge or falling edge trigger.
Figure 5-17:
Press the soft keys corresponding with trigger setting
8. Press the Graph key to view the waveform parameters in the graph mode.
To set up frequency sweep from a remote interface:
SWEep:SPACing {LINear|LOGarithmic}
SWEep:TIME {<seconds>|MINimum|MAXimum}
FREQuency:STARt {<frequency>|MINimum|MAXimum}
FREQuency:STOP {<frequency>|MINimum|MAXimum}
FREQuency:CENTer {<frequency>|MINimum|MAXimum}