User manual
3390-900-01 Rev. C / January 2009 Return to Section Topics B-7
Model 3390 Arbitrary Waveform Generator User’s Manual Appendix B: Error Messages
-222 Data out of range; pulse edge
time limited by duty cycle; value
clipped to...
The specified pulse edge time does not conform to
the following condition, and is adjusted to
accommodate the duty cycle: Edge Time
≤
0.625 x
Period x Duty Cycle.
-222 Data out of range; period;
value clipped to...
The pulse period is out of range and is adjusted.
-222 Data out of range; frequency;
value clipped to...
The frequency is out of range and is adjusted.
-222 Data out of range; user
frequency; value clipped to upper
limit
The frequency is out of range and is adjusted to the
maximum value allowed to accommodate the
selected arbitrary waveform.
-222 Data out of range; ramp
frequency; value clipped to upper
limit
The frequency is out of range and is adjusted to the
maximum value allowed to accommodate the
selected ramp waveform.
-222 Data out of range; pulse
frequency; value clipped to upper
limit
The frequency is out of range and is adjusted to the
maximum value allowed to accommodate the
selected pulse waveform.
-222 Data out of range; burst
period; value clipped to...
The burst period is out of range and is adjusted.
-222 Data out of range; burst
count; value clipped to...
The burst count is out of range and is adjusted.
-222 Data out of range; burst
period limited by length of burst;
value clipped to lower limit
The burst period is too short to accommodate the
burst count and frequency, and is adjusted to the
minimum value allowed.
-222 Data out of range; burst count
limited by length of burst; value
clipped to upper limit
The burst count is out of range. When the immediate
trigger source is selected, the burst count must be
less than the product of the burst period and the
waveform frequency.
-222 Data out of range; amplitude;
value clipped to...
The amplitude is out of range and is adjusted.
-222 Data out of range; offset;
value clipped to...
The offset voltage is out of range and is adjusted.
-222 Data out of range; frequency
in burst mode; value clipped to...
The frequency is adjusted, as indicated by the burst
period.
-222 Data out of range; frequency
in FM; value clipped to...
The carrier frequency is adjusted to the lower
boundary determined by the
FM:DEV command.
The carrier frequency must be greater than or equal
to the frequency deviation.
-222 Data out of range; marker
confined to sweep span; value
clipped to...
The specified marker frequency is out of range and is
adjusted to the start or stop frequency, whichever is
closer.
-222 Data out of range; pulse
width; value clipped to...
The desired pulse width is limited to the upper or
lower boundary, as indicated by the instrument
hardware.
-222 Data out of range; pulse edge
time; value clipped to...
The desired edge time is limited to the upper or lower
boundary, as indicated by the instrument hardware.
-222 Data out of range; FM
deviation; value clipped to...
The desired frequency deviation is limited to the
upper or lower boundary set by the carrier frequency.
-222 Data out of range; FM
deviation limited by minimum
frequency
The frequency deviation is limited to the lower limit.
Table B-2:
Execution errors
Error code Description