-221 Settings conflict; frequency
changed for pulse function
The maximum output frequency for a pulse
waveform is 10 MHz. When you change function
from one allowing higher frequency to pulse
waveform, the frequency is adjusted to 10 MHz.
-221 Settings conflict; frequency
reduced for ramp function
The maximum output frequency for a ramp waveform
is 200 kHz. When you change function from one
allowing higher frequency to ramp waveform, the
frequency is adjusted to 200 kHz.
-221 Settings conflict; frequency
made compatible with burst mode
The minimum output frequency for an internally-
triggered burst is 2.001 mHz. The waveform
generator has adjusted the frequency to be
compatible with the current setting.
-221 Settings conflict; burst
turned off by selection of other
mode or modulation
Only one modulation can be enabled at one time.
-221 Settings conflict; FSK turned
off by selection of other mode or
modulation
Only one modulation can be enabled at one time.
-221 Settings conflict; FM turned
off by selection of other mode or
modulation
Only one modulation can be enabled at one time.
-221 Settings conflict; AM turned
off by selection of other mode or
modulation
Only one modulation can be enabled at one time.
-221 Settings conflict; PM turned
off by selection of other mode or
modulation
Only one modulation can be enabled at one time.
-221 Settings conflict; PWM turned
off by selection of other mode or
modulation
Only one modulation can be enabled at one time.
-221 Settings conflict; sweep
turned off by selection of other
mode or modulation
Only one modulation can be enabled at one time.
-221 Settings conflict; not able to
modulate this function
This waveform generator cannot generate an AM,
FM, PM, or FSK modulated waveform using the
pulse, noise, or DC voltage function.
-221 Settings conflict; PWM only
available in pulse function
The waveform generator can only generate a PWM
modulated waveform using the pulse function.
-221 Settings conflict; not able to
sweep this function
The waveform generator cannot generate a sweep
using the pulse, noise, or DC voltage function.
-221 Settings conflict; not able to
burst this function
The waveform generator cannot generate a burst
using the DC voltage function.
-221 Settings conflict; not able to
modulate noise, modulation turned
off
The waveform generator cannot generate a
modulated waveform using the noise function.
-221 Settings conflict; not able to
sweep pulse, sweep turned off
The waveform generator cannot generate a sweep
using the pulse function.
-221 Settings conflict; not able to
modulate dc, modulation turned off
The waveform generator cannot generate a
modulated DC voltage function.
-221 Settings conflict; not able to
sweep dc, sweep turned off
The waveform generator cannot generate a sweep
using the DC voltage function.