Operation Manual
F-4 Temperature Equations Model 2750 Multimeter/Switch System User’s Manual
Table F-4
Type K inverse function polynomial
-200°C to 0°C
(-5,891µV to 0µV)
0°C to 500°C
(0µV to 20,644µV)
500°C to 1,372°C
(20,644µV to 54,886µV)
c
0
 =  0.0  0.0 -1.318 058 ×  10
2
c
1
 =  2.517 346 2 ×  10
-2
 2.508 355 2 ×  10
-2
 4.830 222 ×  10
-2
c
2
 = -1.166 287 8 ×  10
-6
 7.860 106 2 ×  10
-8
-1.646 031 × 10
-6
c
3
 = -1.083 363 8 ×  10
-9
-2.503 131 2 ×  10
-10
 5.464 731 ×  10
-11
c
4
 = -8.977 354 0 ×  10
-13
 8.315 270 2 ×  10
-14
-9.650 715 ×  10
-16
c
5
 = -3.734 237 7 ×  10
-16
-1.228 034 2 ×  10
-17
 8.802 193 ×  10
-21
c
6
 = -8.663 264 3 ×  10
-20
 9.804 036 2 ×  10
-22
-3.110 810 ×  10
-26
c
7
 = -1.045 059 8 ×  10
-23
-4.413 030 2 ×  10
-26
c
8
 = -5.192 057 7 ×  10
-28
 1.057 734 2 ×  10
-30
c
9
 = -1.052 755 2 ×  10
-35
Error:   0.04°C to -0.02°C  0.04°C to -0.05°C  0.06°C to -0.05°C
t
90
 = c
0
 + c
1
E + c
2
E
2
 + c
3
E
3
 ... c
i
E
i
where: t
90
 is the calculated temperature in °C.
E is the measured voltage in microvolts.
Table F-5
Type N inverse function polynomial
-200°C to 0°C
(-3,990µV to 0µV)
0°C to 600°C
(0uV to 20,613µV)
600°C to 1,300°C
(20,613µV to 47,513µV)
c
0
 =  0.0  0.0  1.972 485 ×  10
1
c
1
 =  3.843 684 7 ×  10
-2
 3.868 96 ×  10
-2
 3.300 943 ×  10
-2
c
2
 =  1.101 048 5 ×  10
-6
-1.082 67 ×  10
-6
-3.915 159 ×  10
-7
c
3
 =  5.222 931 2 ×  10
-9
 4.702 05 ×  10
-11
 9.855 391 ×  10
-12
c
4
 =  7.206 052 5 ×  10
-12
-2.121 69 ×  10
-18
-1.274 371 ×  10
-16
c
5
 =  5.848 858 6 ×  10
-15
-1.172 72 ×  10
-19
 7.767 022 ×  10
-22
c
6
 =  2.775 491 6 ×  10
-18
 5.392 80 ×  10
-24
c
7
 =  7.707 516 6 ×  10
-22
-7.981 56 ×  10
-29
c
8
 =  1.158 266 5 ×  10
-25
c
9
 =  7.313 886 8 ×  10
-30
Error:   0.03°C to -0.02°C  0.03°C to -0.02°C  0.02°C to -0.04°C
t
90
 = c
0
 + c
1
E + c
2
E
2
 + c
3
E
3
 ... c
i
E
i
where: t
90
 is the calculated temperature in °C.
E is the measured voltage in microvolts.
2750-900-01.book Page 4 Wednesday, August 3, 2011 7:56 AM