Operation Manual
7-24 Scanning Model 2750 Multimeter/Switch System User’s Manual
Monitor scan (analog trigger)
A channel can be assigned as a monitor channel. When the monitor channel detects that a 
reading limit has been reached, the scan will be triggered to start. 
 There are four reading limits that can be used to trigger the start of the scan: low limit 1 
(LLIM1), high limit 1 (HLIM1), low limit 2 (LLIM2), and high limit 2 (HLIM2). The scan 
will start when any enabled reading limit event is detected by the monitor channel. Details 
on Limits are provided in Section 9.
NOTE An overflow reading (“OVRFLW” message displayed) is interpreted by the 
Model 2750 as a positive reading, even if the input signal is negative. This could 
inadvertently trigger a monitor scan. For example, assume the monitor channel 
is monitoring a negative input signal, and the instrument is configured to trigger 
a monitor scan if a positive input signal is detected. If for some reason, the neg-
ative input signal exceeds the measurement range, the overflow reading will be 
interpreted as positive and trigger the start of the scan. 
Perform the following steps to run a monitor scan:
NOTE The last enabled scan function (STEP or SCAN) will be used for the monitor 
scan.
1. Perform Step 1 and Step 2 of the “Advanced scan setup procedure” to set up scan 
channels. With the channel to be used as the monitor selected, set and enable limits 
as follows. Note that you only need to set values for limits that are going to be 
used.
a. Press SHIFT and then LIMITS to access the limits menu. Note that the CHAN 
annunciator flashes to indicate that the menu is being used to set up a scan 
channel.
b. Use the , , 
Δ , and ∇ keys to set high limit 1 (HI1) and press ENTER.
c. Set low limit 1 (LO1) and press ENTER.
d. Set high limit 2 (HI2) and press ENTER.
e. Set low limit 2 (LO2) and press ENTER. The instrument returns to the scan 
setup menu.
f. Press SHIFT and then ON/OFF to display the present state of LIMITS (ON or 
OFF). Again, the CHAN annunciator flashes to indicate that the menu is for a 
scan channel.
g. Press the 
Δ  or ∇ key to display “LIMITS:ON” and press ENTER. The instru-
ment returns to the scan setup menu. Note that the HIGH and LOW annuncia-
tors are on to indicate that limits are enabled.
After all scan channels are set up, press ENTER. The present state of IMM SCAN 
is Y (yes) or N (no).
2. Press the 
Δ  or ∇ key to display IMM SCAN: N and press ENTER.
2750-900-01.book Page 24 Wednesday, August 3, 2011 7:56 AM