Model 2750 Multimeter/Switch System User’s Manual Scanning 7-9
Delay (Auto or Manual) — The user can select either auto delay or manual delay. With 
auto delay selected, the instrument automatically selects a delay period that will provide 
sufficient settling for function and autorange changes, and multi-phase measurements.
The auto delay period cannot be adjusted by the user. It is a fixed delay for the selected 
function and range (Table 8-1).
NOTE When scanning, the auto delay times in Table 8-1 are valid for all control 
sources (Immediate, External, Timer, Manual, or Bus).
With manual delay selected, the user can set the delay period from 0 seconds to 99 hours, 
99 minutes, 99.999 seconds. However, if you set a delay shorter than the corresponding 
auto delay period, measurement uncertainty increases (noisy and/or unsettled readings 
may result).
NOTE Keep in mind that if the timer control source is selected, the Delay period is only 
in effect for the first pass through the loop. 
Ratio/Chan Average Delay — With ratio or channel average enabled, a delay is typically 
used to keep the channel relays from cycling too fast. The default delay period is 0.5 sec-
onds but can be set from 0 to 999999.999 seconds using remote programming (see 
Section 5 for details). Ratio and channel average are covered in Section 5. 
NOTE The Ratio/Chan Average Delay is in addition to the Timer or Delay (Auto or 
Manual). That is, it occurs after the Timer interval or Delay period elapses. 
Device action
The channel measurement process is performed at this block. If repeat filter is enabled, the 
filter process is also performed.
Reading count
NOTE For both STEP and SCAN, the reading count specifies the number of readings to 
store in the buffer.
STEP operation — The reading count specifies the number of channels to scan. This can 
be equal to, less than, or greater than the number of channels in the scan list. The last 
scanned channel remains closed. If you start the scan again, it will start at the next 
channel.
If the reading count is set to infinity (INF), the scan will continuously repeat until you 
stop it.
NOTE One counter is used for STEP operation. As shown in Figure 7-1, reading count 
sets the Trigger Counter.