Operation Manual
3-42 Basic DMM Operation Model 2750 Multimeter/Switch System User’s Manual
Remote programming for basic measurements
Basic measurement commands
NOTE When measurements are performed, the readings are fed to other enabled pro-
cessing operations. Appendix D explains “Data flow (remote operation)” and 
the commands used to return the various processed readings. 
Commands to perform basic measurements are listed in Table 3-8.
Table 3-8
Basic measurement commands
Commands
1
Description Default Ref
Select measurement function
[SENSe[1]] Optional root command.
:FUNCtion <name> [, <clist>] Select measurement function: 
<name> = ‘VOLTage[:DC]’,
‘VOLTage:AC’, ‘CURRent[:DC]’, 
‘CURRent:AC’, ‘RESistance’, 
‘FRESistance’, ‘CONTinuity’, 
‘FREQuency’, or ‘PERiod’.
Note: [:DC] is optional.
VOLT a
DCV function
[SENSe[1]] Optional root command.
:VOLTage[:DC]:IDIVider <b> Enable/disable DCV input divider; 
<b> = ON or OFF. 
OFF
Ω
4 function
[SENSe[1]] Optional root command.
:FRESistance:OCOMpensated <b> 
[, <clist>]
Enable/disable offset compensated ohms.  OFF b
:FRESistance:DCIRcuit <b> [, <clist>] Enable/disable dry circuit ohms.  OFF b
2750-900-01.book Page 42 Wednesday, August 3, 2011 7:56 AM