Volts measurement procedure
NOTE Make sure the INPUTS switch is in the correct position. To use front panel 
inputs, it must be in the “F” (out) position. For switching modules, it must be in 
the “R” (in) position.
1. If a switching channel is presently closed (displayed), press OPEN to open it.
2. Select the volts measurement function by pressing DCV or ACV.
3. Use the RANGE 
Δ  and ∇ keys to select a measurement range consistent with the 
expected voltage, or press AUTO to select autoranging (AUTO annunciator turns 
on). Details on range are provided in Section 4.
4. Apply the voltage(s) to be measured.
CAUTION Do not apply more than maximum input levels indicated in 
Figure 3-2 
and Figure 3-3 or instrument damage may occur. The voltage limit is 
subject to the 8 ×  10
7
VHz product.
Model 7700 switching module — The maximum allowable voltage is 
300V DC or RMS. Exceeding these limits may cause damage to the 
switching module.
WARNING If both the front panel terminals and the switching module terminals 
are connected at the same time, the test leads must be rated to the high-
est voltage that is connected. For example, if 1000V is connected to the 
front panel input, the test lead insulation for the switching module 
must also be rated for 1000V.
5. If using a switching module, perform the following steps to close the desired 
channel: 
a. Press the CLOSE key. 
b. Use , , 
Δ , and ∇ to key in the channel number and press ENTER. The 
previously closed channel (if there is one) will open, and the specified channel 
will close.
NOTE While in the normal measurement state, you can use the  and  keys to close 
channels. In general, each key press will open the presently closed channel, and 
then close the next higher or lower channel.
6. Observe the displayed reading. If the “OVERFLOW” message is displayed, select 
a higher range until a normal reading is displayed (or press AUTO for autorang-
ing). For manual ranging, use the lowest possible range for the best resolution.
7. To measure other switching channels, repeat steps 5 and 6. 
8. When finished, press OPEN if there is a channel closed. 
2750-900-01.book Page 11 Wednesday, August 3, 2011 7:56 AM