Datasheet
DIGITAL MULTIMETERS & SYSTEMS
A Greater Measure of Confidence
www.keithley.com
1.888.KEITHLEY
(U.S. only)
DC CHARACTERISTICS
1
Conditions: MED (1 PLC)
2
or 10 PLC or MED (1 PLC) with Digital Filter of 10
Test Current
±5% or
Burden
Voltage
Input Resistance or
Open Circuit Voltage
3
Accuracy: ±(ppm of reading + ppm of range) (ppm
= parts per million) (e.g., 10ppm = 0.001%)
Temperature
Coefficient
0°–18°C & 28°–50°C
2
4 Hour
4
23°C ±1°
90 Day
23°C ±5°
1 Year
23°C ±5°Function Range Resolution 2700/2701 2750
Voltage
11
100.0000 mV 0.1 µV
>10 GW >10 GW
15 + 30 25 + 35 30 + 35 (1 + 5)/°C
1.000000 V 1.0 µV
>10 GW >10 GW
15 + 6 25 + 7 30 + 7 (1 + 1)/°C
10.00000 V 10 µV
>10 GW >10 GW
10 + 4 20 + 5 30 + 5 (1 + 1)/°C
100.0000 V 100 µV
10 MW ±1% 10 MW ±1%
15 + 6 35 + 9 45 + 9 (5 + 1)/°C
1000.000 V
5
1 mV
10 MW ±1% 10 MW ±1%
20 + 6 35 + 9 50 + 9 (5 + 1)/°C
Resistance
6, 8
1.000000 W
24
1 µW
10 mA 5.9 V 80 + 40 80 + 40 100 + 40 (8 + 1)/°C
10.00000 W
24
10 µW
10 mA 5.9 V 20 + 20 80 + 20 100 + 20 (8 + 1)/°C
100.0000 W 100 µW
1 mA 6.9 V 12.2 V 20 + 20 80 + 20 100 + 20 (8 + 1)/°C
1.000000 kW 1 mW
1 mA 6.9 V 12.2 V 20 + 6 80 + 6 100 + 6 (8 + 1)/°C
10.00000 kW 10 mW
100 µA 6.9 V 6.8 V 20 + 6 80 + 6 100 + 6 (8 + 1)/°C
100.0000 kW 100 mW
10 µA 12.8 V 12.8 V 20 + 6 80 + 10 100 + 10 (8 + 1)/°C
1.000000 MW
23
1.0 W
10 µA 12.8 V 12.8 V 20 + 6 80 + 10 100 + 10 (8 + 1)/°C
10.00000 MW
7, 23
10 W 0.7 µA//10 MW
7.0 V 7.0 V 150 + 6 200 + 10 400 + 10 (70 + 1)/°C
100.0000 MW
7, 23
100 W 0.7 µA//10 MW
7.0 V 7.0 V 800 + 30 2000 + 30 2000 + 30 (385 + 1)/°C
Dry Circuit
Resistance
21, 24
1.000000 W 1 µW
10 mA 20 mV 80 + 40 80 + 40 100 + 40 (8 + 1)/°C
10.00000 W 10 µW
1 mA 20 mV 25 + 40 80 + 40 100 + 40 (8 + 1)/°C
100.0000 W 100 µW
100 µA 20 mV 25 + 40 90 + 40 140 + 40 (8 + 1)/°C
1.000000 kW 1 mW
10 µA 20 mV 25 + 90 180 + 90 400 + 90 (8 + 1)/°C
Continuity (2W)
1.000 kW 100 mW
1 mA 6.9 V 12.2 V 40 + 100 100 + 100 100 + 100 (8 + 1)/°C
Current
20.00000 mA 10 nA < 0.2 V 60 + 30 300 + 80 500 + 80 (50 + 5)/°C
100.0000 mA 100 nA < 0.1 V 100 + 300 300 + 800 500 + 800 (50 + 50)/°C
1.000000 A 1.0 µA < 0.5 V
9
200 + 30 500 + 80 800 + 80 (50 + 5)/°C
3.000000 A 10 µA < 1.5 V
9
1000 + 15 1200 + 40 1200 + 40 (50 + 5)/°C
Channel (Ratio)
10
Ratio Accuracy = Accuracy of selected Channel Range + Accuracy of Paired Channel Range
Channel (Average)
10
Average Accuracy = Accuracy of selected Channel Range + Accuracy of Paired Channel Range
TEMPERATURE
19
(Displayed in °C, °F, or K. Exclusive of probe errors.)
Thermocouples (Accuracy based on ITS-90)
90 Day/1 Year
(23°C ± 5°C)
T
emperature
Coefficient
0°–18°C & 28°–50°CTy
pe Range Resolution
Relative to
Simulated
Reference Junction
Using
77XX
Module*
J –200 to + 760°C 0.001 °C 0.2°C 1.0°C 0.03°C/°C
K –200 to +1372°C 0.001 °C 0.2°C 1.0°C 0.03°C/°C
N –200 to +1300°C 0.001 °C 0.2°C 1.0°C 0.03°C/°C
T –200 to + 400°C 0.001 °C 0.2°C 1.0°C 0.03°C/°C
E –200 to +1000°C 0.001 °C 0.2°C 1.0°C 0.03°C/°C
R 0 to +1768°C 0.1 °C 0.6°C 1.8°C 0.03°C/°C
S 0 to +1768°C 0.1 °C 0.6°C 1.8°C 0.03°C/°C
B +350 to +1820°C 0.1 °C 0.6°C 1.8°C 0.03°C/°C
* Using 7710 Module: J: 2.5°C; K: 1°C. N, T, E Types: 1.5°C. R, S, B Types: 2.7°C.
4-Wire RTD:
(100W
platinum [PT100], D100, F100, PT385, PT3916, or user type. Offset compensation On.)
–200° to +630°C 0.01°C 0.06°C 0.003°C/°C
Thermistor: (2.2kW, 5kW, and 10kW)
20
–80° to +150°C 0.01°C 0.08°C 0.002°C/°C
DC SYSTEM SPEEDS
15,18
2700/2750 2701
RANGE CHANGES (excludes 4WW)
16
:
50/s (42/s) 50/s (42/s)
FUNCTION CHANGES
16
:
50/s (42/s) 50/s (42/s)
AUTORANGE TIME
16
:
<30 ms <30 ms
ASCII READINGS TO RS-232 (19.2k baud):
55/s 300/s
MAX. EXTERNAL TRIGGER RATE:
375/s 2000/s
DC MEASUREMENT SPEEDS
15
Single Channel, 60Hz (50Hz) Operation
Function Digits Readings/s PLCs
DCV, DCI, W (<10M),
Thermocouple,
Thermistor
6.5
12,16
5 (4) 10
6.5
16
35 (28) 1
6.5
12,16
45 (36) 1
5.5
12,16
150 (120) 0.1
5.5
16, 17
300 (240) 0.1
5.5
17
500 (400) 0.1
2700 and 2750 only
4.5
17
2500 (2000) 0.01
2701 only
3.5 3500 (3500) 0.002
4WW (<10M)
6.5
16
1.4 (1.1) 10
6.5
16
15 (12) 1
5.5
17
33 (25) 0.1
4WW OComp, RTD
22
6.5
16
0.9 (0.7) 10
6.5
16
8 (6.4) 1
5.5
16, 17
18 (14.4) 0.1
Channel (Ratio),
Channel (AVG)
6.5
16
2.5 (2) 10
6.5
16
15 (12) 1
5.5
17
25 (20) 0.1
Multiple Channels, Into Memory
18
Channels/s
2700 2701 2750
7710 Scanning DCV 180/s 500/s 230/s
7710 Scanning DCV with Limits or Time Stamp On 170/s 500/s 230/s
7710 Scanning DCV alternating 2WW
45/s 115/s 60/s
Multiple Channels, Into and Out of
Memory to GPIB
16, 18
or Ethernet
Channels/s
2700 2701 2750
7702 Scanning DCV 65/s 75/s 65/s
7700 and 7708 Scanning Temperature (T/C) 50/s 50/s 50/s
7710 Scanning DCV 145/s 440/s 210/s
7710 Scanning DCV with Limits or Time Stamp On 145/s 440/s 210/s
7710 Scanning DCV alternating 2WW
40/s 115/s 55/s
2700, 2701,
2750
Multimeter/Data Acquisition/
Switch Systems
Integra Series condensed specifications
Integra Series condensed specifications