Datasheet

TEK.COM 7
2600B System SourceMeter
®
SMU Instruments
Current Source Specifications
Current Programming Accuracy
Range
Programming
Resolution
Accuracy (1 Year), 23°C ±5°C
±(% rdg. + amps)
Typical Noise (peak-peak)
0.1Hz–10Hz
100 nA 2 pA 0.06% + 100 pA 5 pA
1 µA 20 pA 0.03% + 800 pA 25 pA
10 µA 200 pA 0.03% + 5 nA 60 pA
100 µA 2 nA 0.03% + 60 nA 3 nA
1 mA 20 nA 0.03% + 300 nA 6 nA
10 mA 200 nA 0.03% + 6 µA 200 nA
100 mA 2 µA 0.03% + 30 µA 600 nA
1 A
1
20 µA 0.05% + 1.8 mA 70 µA
3 A
1
20 µA 0.06% + 4 mA 150 µA
10 A
1, 2
200 µA 0.5% + 40 mA (typical)
Temperature Coefficient (0°–18°C and 28°50°C)
3
±(0.15 × accuracy specification)/°C.
Maximum Output Power and Source/Sink Limits
4
40.4 W per channel maximum. ±1.01 A @ ±40.0 V, ±3.03 A @ ±6.0 V, four quadrant source or
sink operation.
Current Regulation Line: 0.01% of range. Load: ±(0.01% of range + 100 pA).
Voltage Limit/Compliance
5
Bipolar voltage limit (compliance) set with a single value. Minimum value is 10 mV. Accuracy is the same as
voltage source.
Overshoot <±0.1% typical (step size = 10% to 90% of range, resistive load; see Current Source Output Settling Time
for additional test conditions).
NOTES
1. Full power source operation regardless of load to 30°C ambient. Above 30°C and/or power sink operation, refer to “Operating Boundaries” in the Series 2600B Reference Manual for
additional power derating information.
2. 10 A range accessible only in pulse mode.
3. High Capacitance Mode accuracy is applicable at 23°C ±5°C only.
4. Full power source operation regardless of load to 30°C ambient. Above 30°C and/or power sink operation, refer to “Operating Boundaries” in the Series 2600B Reference Manual for
additional power derating information.
5. For sink mode operation (quadrants II and IV), add 10% of compliance range and ±0.02% of limit setting to corresponding voltage source specification. For 100 mV range add an additional
60 mV of uncertainty.
Additional Source Specifications
Transient Response Time <70 µs for the output to recover to within 0.1% for a 10% to 90% step change in load.
Voltage Source Output Settling Time
Time required to reach within 0.1% of final value after source level command is processed on a fixed range.
100 mV, 1 V Ranges <50 µs typical.
6 V Range <100 µs typical.
40 V Range
1
<150 µs typical.
NOTES
1. Add 150 µs when measuring on the 1 A range.