Datasheet
Datasheet
TEK.COM6
Specification Conditions (2601B, 2602B, 2604B)
This document contains specifications and supplemental information for the Models 2601B, 2602B, and 2604B System SourceMeter
®
SMU
instruments. Specifications are the standards against which the Models 2601B, 2602B, and 2604B are tested. Upon leaving the factory, the
2601B, 2602B, and 2604B meet these specifications. Supplemental and typical values are non- warranted, apply at 23°C, and are provided solely
as useful information.
Accuracy specifications are applicable for both normal and high capacitance modes.
The source and measurement accuracies are specified at the SourceMeter CHANNEL A (2601B, 2602B, and 2604B) or SourceMeter
CHANNEL B (2602B and 2604B) terminals under the following conditions:
1. 23°C ± 5°C, <70% relative humidity
2. After 2 hour warm-up
3. Speed normal (1 NPLC)
4. A/D auto-zero enabled
5. Remote sense operation or properly zeroed local operation
6. Calibration period = 1 year
Source Specifications (2601B, 2602B, 2604B)
Voltage Source Specifications
Voltage Programming Accuracy
1
Range
Programming
Resolution
Accuracy (1 Year), 23°C ±5°C
±(% rdg. + volts)
Typical Noise (peak-peak)
0.1Hz–10Hz
100 mV 5 µV 0.02% + 250 µV 20 µV
1 V 50 µV 0.02% + 400 µV 50 µV
6 V 50 µV 0.02% + 1.8 mV 100 µV
40 V 500 µV 0.02% + 12 mV 500 µV
Temperature Coefficient (0°–18°C and 28°–50°C)
2
±(0.15 × accuracy specification)/°C. Applicable for normal mode only. Not applicable for high
capacitance mode.
Maximum Output Power and Source/Sink Limits
3
40.4 W per channel maximum. ±40.4 V @ ±1.0 A, ±6.06 V @ ±3.0 A, four quadrant source or
sink operation.
Voltage Regulation Line: 0.01% of range. Load: ±(0.01% of range + 100 µV).
Noise 10 Hz–20 MHz <20 mV peak-peak (typical), <3 mV RMS (typical), 6 V range.
Current Limit/Compliance
4
Bipolar current limit (compliance) set with single value. Minimum value is 10 nA. Accuracy same as
current source.
Overshoot <±(0.1% + 10 mV) typical. Step size = 10% to 90% of range, resistive load, maximum current
limit/compliance.
Guard Offset Voltage <4 mV typical. Current <10 mA.
NOTES
1. Add 50 µV to source accuracy specifications per volt of HI lead drop.
2. High Capacitance Mode accuracy is applicable at 23°C ±5°C only.
3. Full power source operation regardless of load to 30°C ambient. Above 30°C and/or power sink operation, refer to “Operating Boundaries” in the Series 2600B Reference Manual for
additional power derating information.
4. For sink mode operation (quadrants II and IV), add 0.06% of limit range to the corresponding current limit accuracy specifications. Specifications apply with sink mode operation enabled.