Datasheet
TEK.COM 29
2600B System SourceMeter
®
SMU Instruments
Safety Interlock Pin Active high input. >3.4 V @ 24 mA (absolute maximum of 6 V) must be externally applied to this pin
to ensure 200 V operation. This signal is pulled down to chassis ground with a 10 kΩ resistor. 200
V operation will be blocked when the INTERLOCK signal is <0.4 V (absolute minimum –0.4 V). See
figure below:
Read by firmware
+220 V Supply
–220 V Supply
INTERLOCK Pin
(on DIGITAL I/O
connector)
Rear Panel
Chassis
Ground
To output stage
10kΩ
Coil
Resistance
145Ω ±10 %
USB File System (Front) USB 2.0 Host: Mass storage class device.
Power Supply 100 V to 250 VAC, 50–60 Hz (auto sensing), 240 VA max.
Cooling Forced air. Side intake and rear exhaust. One side must be unobstructed when rack mounted.
EMC Conforms to European Union Directive 2004/108/EEC, EN 61326-1.
Safety Conforms to European Union Directive 73/23/EEC, EN 61010-1, and UL 61010-1.
Dimensions 89 mm high × 213 mm wide × 460 mm deep (3½ in × 8
3
⁄8 in × 17½ in).
Bench Configuration (with handle and feet): 104mm high × 238mm wide × 460mm deep
(4
1
⁄8 in × 9
3
⁄8 in × 17½ in).
Weight 2635B: 4.75 kg (10.4 lbs). 2634B, 2636B: 5.50 kg (12.0 lbs).
Environment For indoor use only.
Altitude Maximum 2000 meters above sea level.
Operating 0°–50°C, 70% R.H. up to 35°C. Derate 3% R.H./°C, 35°–50°C.
Storage –25°C to 65°C.
See pages 30 and 31 for measurement speeds and other specifications.