Datasheet

TEK.COM 25
2600B System SourceMeter
®
SMU Instruments
Quadrant Diagram
+1.5A
+10A
–10A
–1A
+1A
+0.1A
–0.1A
–1.5A
+200V+5V
–5V
+20V–20V 0V
0A
+180V–180V–200V
DC
Pulse
Pulse
Pulse
2
2
2
2
4
4
1
3
3
3
3
NOTES
1. Times measured from the start of pulse to the start off-time; see figure below.
Pulse Level
Bias Level
Start t
on
Start t
off
90%
10%
t
on
t
off
10%
2. Thermally limited in sink mode (quadrants II and IV) and ambient temperatures above 30°C. See power equations in the reference manual for more information.
3. Voltage source operation with 1.5 A current limit.
4. Typical performance for minimum settled pulse widths:
Source Value Load Source Settling (% of range) Min. Pulse Width
5 V 0.5 Ω 1% 300 µs
20 V 200 Ω 0.2% 200 µs
180 V 180 Ω 0.2% 5 ms
200 V (1.5 A Limit) 200 Ω 0.2% 1.5 ms
100 mA 200 Ω 1% 200 µs
1 A 200 Ω 1% 500 µs
1 A 180 Ω 0.2% 5 ms
10 A 0.5 Ω 0.5% 300 µs
Typical tests were performed using remote operation, 4W sense, and best, fixed measurement range. For more information on pulse scripts, see the Series 2600B Reference Manual.