Datasheet
Datasheet
TEK.COM24
Current Source Output Settling Time
Time required to reach within 0.1% of final value after source level command is processed on a fixed range.
Values below for I
out
× R
load
= 1 V unless noted.
Range Settling Time
1.5 A – 1 A <120 µs (typical) (R
load
> 6Ω)
100 mA – 10 mA <80 µs (typical)
1 mA <100 µs (typical)
100 µA <150 µs (typical)
10 µA <500 µs (typical)
1 µA <2 ms (typical)
100 nA <20 ms (typical)
10 nA <40 ms (typical)
1 nA <150 ms (typical)
DC Floating Voltage Output can be floated up to ±250 VDC from chassis ground.
Remote Sense Operating Range
1
Maximum voltage between HI and SENSE HI = 3 V.
Maximum voltage between LO and SENSE LO = 3 V.
Voltage Output Headroom
200 V Range Max. output voltage = 202.3 V – total voltage drop across source leads (maximum 1 Ω per source lead).
20 V Range Max. output voltage = 23.3 V – total voltage drop across source leads (maximum 1 Ω per source lead).
Over Temperature Protection Internally sensed temperature overload puts unit in standby mode.
Voltage Source Range Change Overshoot
<300 mV + 0.1% of larger range (typical). Overshoot into a 200 kΩ load, 20 MHz BW.
Current Source Range Change Overshoot
<5% of larger range + 300 mV/R
load
(typical with source settling set to SETTLE_SMOOTH_100NA). See
Current Source Output Settling Time for additional test conditions.
NOTES
1. Add 50 µV to source accuracy specifications per volt of HI lead drop.
Pulse Specifications
Region
Maximum
Current Limit
Maximum
Pulse Width
1
Maximum
Duty Cycle
2
1 100 mA @ 200 V DC, no limit 100%
1 1.5 A @ 20 V DC, no limit 100%
2 1 A @ 180 V 8.5 ms 1%
3
3
1 A @ 200 V 2.2 ms 1%
4 10 A @ 5 V 1 ms 2.2%
Minimum Programmable 100 µs.
Pulse Width
4, 5
Note: Minimum pulse width for settled source at a given I/V output and load can be longer than 100 µs.
Pulse Width
Programming Resolution 1 µs.
Pulse Width
Programming Accuracy
5
±5 µs.
Pulse Width Jitter 50 µs (typical).