Datasheet

TEK.COM 13
2600B System SourceMeter
®
SMU Instruments
USB File System (Front) USB 2.0 Host: Mass storage class device.
Power Supply 100 V to 250 VAC, 5060 Hz (auto sensing), 240 VA max.
Cooling Forced air. Side intake and rear exhaust. One side must be unobstructed when rack mounted.
EMC Conforms to European Union Directive 2004/108/EEC, EN 61326-1.
Safety Conforms to European Union Directive 73/23/EEC, EN 61010-1, and UL 61010-1.
Dimensions 89 mm high × 213 mm wide × 460 mm deep (3½ in × 8
3
8 in × 17½ in).
Bench Configuration (with handle and feet): 104mm high × 238mm wide × 460mm deep
(4
1
8 in × 9
3
8 in × 17½ in).
Weight 2601B: 4.75 kg (10.4 lbs). 2602B, 2604B: 5.50 kg (12.0 lbs).
Environment For indoor use only.
Altitude Maximum 2000 meters above sea level.
Operating 0°–50°C, 70% R.H. up to 35°C. Derate 3% R.H./°C, 35°–50°C.
Storage 25°C to 65°C.
See pages 30 and 31 for measurement speeds and other specifications.