Datasheet

Datasheet
TEK.COM10
Meter Specifications (2601B, 2602B, 2604B)
Voltage Measurement Accuracy
1, 2
Range Default Display Resolution
3
Input Resistance
Accuracy (1 Year), 23°C ±5°C
±(% rdg. + volts)
100 mV 100 nV >10 G Ω 0.015% + 150 µV
1 V 1 µV >10 G Ω 0.015% + 200 µV
6 V 10 µV >10 G Ω 0.015% + 1 mV
40 V 10 µV >10 G Ω 0.015% + 8 mV
Temperature Coefficient (0°–18°C and 28°50°C)
4
±(0.15 × accuracy specification)/°C. Applicable for normal mode only. Not applicable for high
capacitance mode.
Current Measurement Accuracy
2
Range Default Display Resolution
5
Voltage Burden
6
Accuracy (1 Year), 23°C ±5°C
±(% rdg. + amps)
100 nA 100 fA <1 mV 0.05% + 100 pA
1 µA 1 pA <1 mV 0.025% + 500 pA
10 µA 10 pA <1 mV 0.025% + 1.5 nA
100 µA 100 pA <1 mV 0.02% + 25 nA
1 mA 1 nA <1 mV 0.02% + 200 nA
10 mA 10 nA <1 mV 0.02% + 2.5 µA
100 mA 100 nA <1 mV 0.02% + 20 µA
1 A 1 µA <1 mV 0.03% + 1.5 mA
3 A 1 µA <1 mV 0.05% + 3.5 mA
10 A
7
10 µA <1 mV 0.4% + 25 mA (typical)
Current Measure Settling Time (time for measurement to settle after a V
step
)
8
Time required to reach within 0.1% of final value after source level command is processed on a fixed range.
Values for V
out
= 1 V unless noted.
Current Range 1 mA.
Settling Time <100 µs (typical).
Temperature Coefficient (0°–18°C and 28°50°C)
9
±(0.15 × accuracy specification/°C. Applicable for normal mode only. Not applicable for high
capacitance mode.
NOTES
1. Add 50µV to source accuracy specifications per volt of HI lead drop.
2. e-rate accuracy specifications for NPLC setting <1 by increasing error term.
Add appropriate % of range term using table below.
NPLC Setting 100 mV Range 1 V–40 V Ranges 100 nA Range 1 µA–100 mA Ranges 1 A3 A Ranges
0.1 0.01% 0.01% 0.01% 0.01% 0.01%
0.01 0.08% 0.07% 0.1% 0.05% 0.05%
0.001 0.8 % 0.6 % 1% 0.5 % 1.1 %
3. Applies when in single channel display mode.
4. High Capacitance Mode accuracy is applicable for 23°C ±5°C only.
5. Applies when in single channel display mode.
6. Four-wire remote sense only with current meter mode selected. Voltage measure set to 100mV or 1V range only.
7. 10 A range accessible only in pulse mode.
8. Compliance equal to 100mA.
9. High Capacitance Mode accuracy is applicable for 23°C ±5°C only.