User manual

Series 2600B
System SourceMeter® Instrument Reference Manual Section 7:
TSP command reference
2600BS-901-01 Rev. B / May 2013 7-337
Example
trigger.blender[2].clear()
Clears the event detector for
blender 2.
Also see
None
trigger.blender[N].EVENT_ID
This constant contains the trigger blender event number.
Type TSP-Link accessible
Affected by Where saved Default value
Constant
Yes
Usage
eventID = trigger.blender[N].EVENT_ID
eventID
Trigger event number
N
The blender number (1 to 6)
Details
Set the stimulus of any trigger object to the value of this constant to have the trigger object respond to trigger
events from this trigger blender.
Example
digio.trigger[1].stimulus = trigger.blender[2].EVENT_ID
Set the trigger stimulus of
digital I/O trigger 1 to be
controlled by the trigger
blender 2 event.
Also see
None
trigger.blender[N].orenable
This attribute selects whether the blender operates in OR mode or AND mode.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
Yes
Instrument reset
Trigger blender N reset
Recall setup
Not saved
false (AND mode)
Usage
orenable = trigger.blender[N].orenable
trigger.blender[N].orenable = orenable
orenable
The orenable mode:
true: OR mode
false
: AND mode
N
The trigger blender (1 to 6)