User manual
Series 2600B
System SourceMeterĀ® Instrument Reference Manual Section 7:
TSP command reference
2600BS-901-01 Rev. B / May 2013 7-129
Also see
lan.lxidomain (on page 7-120)
lan.trigger[N].clear() (on page 7-129)
lan.trigger[N].mode (on page 7-133)
lan.trigger[N].overrun (on page 7-134)
lan.trigger[N].stimulus (on page 7-135)
lan.trigger[N].wait() (on page 7-137)
Understanding hardware value and pseudo line state (on page 3-51)
lan.trigger[N].clear()
This function clears the event detector for a trigger.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
lan.trigger[N].clear()
N
The LAN event number to clear (1 to 8)
Details
The trigger event detector enters the detected state when an event is detected. This function clears a trigger
event detector and discards the previous history of the trigger packet.
This function clears all overruns associated with this LAN trigger.
Example
lan.trigger[5].clear()
Clears the event detector with LAN packet 5.
Also see
lan.trigger[N].assert() (on page 7-128)
lan.trigger[N].overrun (on page 7-134)
lan.trigger[N].stimulus (on page 7-135)
lan.trigger[N].wait() (on page 7-137)