Data Sheet
TEK.COM 13
2600B System SourceMeter
®
SMU Instruments
Absolute Maximum Sink Current (flowing into Digital I/O pin)
–11 mA (not including 2604B).
5V Power Supply Pins Limited to 250 mA total for all three pins, solid state fuse protected.
Output Enable Active high input pulled down internally to ground with a 10 kW resistor; when the output enable input
function has been activated, each SourceMeter channel will not turn on unless the output enable pin is
driven to >2.1 V (nominal current = 2.1 V/10 kW = 210 µA).
USB File System (Front) USB 2.0 Host: Mass storage class device.
Power Supply 100 V to 250 VAC, 50–60 Hz (auto sensing), 240 VA max.
Cooling Forced air. Side intake and rear exhaust. One side must be unobstructed when rack mounted.
EMC Conforms to European Union Directive 2004/108/EEC, EN 61326-1.
Safety Conforms to European Union Directive 73/23/EEC, EN 61010-1, and UL 61010-1.
Dimensions 89 mm high × 213 mm wide × 460 mm deep (3½ in × 8
3
⁄8 in × 17½ in).
Bench Configuration (with handle and feet): 104mm high × 238mm wide × 460mm deep
(4
1
⁄8 in × 9
3
⁄8 in × 17½ in).
Weight 2601B: 4.75 kg (10.4 lbs). 2602B, 2604B: 5.50 kg (12.0 lbs).
Environment For indoor use only.
Altitude Maximum 2000 meters above sea level.