User manual

Series 2600B
System SourceMeter® Instrument Reference Manual Appendix B:
Calibration
2600BS-901-01 Rev. B / May 2013 B-37
B. Allow the readings to settle, then get the Series 2600B readings:
r0_hi, r0_lo = smua.contact.r()
C. Characterize both 50 resistors using the resistance function of the digital multimeter.
D. As illustrated in the following figure:
Characterize both 50 Ω resistors using the resistance function of the digital multimeter.
Connect a 50 Ω resistor between the SENSE LO and LO terminals.
Connect the second 50 Ω resistor between the SENSE HI and HI terminals.
Figure 152: Connections for contact check 50 ohm calibration
E. Allow the readings to settle, then get the Series 2600B readings:
r50_hi, r50_lo = smua.contact.r()
F. Send the contact check low calibration adjustment command:
smua.contact.calibratelo(r0_lo, Z_actual, r50_lo, 50_ohm_actual)
Where:
r0_lo
=
Series 2600B 0 low measurement
Z_actual
= Actual zero value; the resistance of the short between the SENSE LO and LO
terminals
r50_lo
=
Series 2600B 50 low measurement
50_ohm_actual
=
Actual 50 resistor value; the actual value of the resistor between the SENSE
LO and LO terminals
Typical values:
smua.contact.calibratelo(r0_lo, 0, r50_lo, 50.15)
Where r0_lo is the same value as measured in step B, and r50_lo is the same value as
measured in step E.